Integrating the Sea-Bird SBE 37 CTD with Aquadopp 2

Updated

Introduction

The Sea-Bird SBE 37 MicroCAT CTD is a conductivity and temperature recorder, with optional pressure measurement, internal memory, serial communication, and an optional pump. Please refer to the Sea-Bird SBE 37 documentation for detailed information on the instrument.

Aquadopp instruments can integrate the SBE 37 as an external sensor through the instrument's 8-pin connector. The CTD receives power from the Aquadopp through the supplied cable.

During a deployment, the CTD measures at the same time as the configured current measurements from the Aquadopp. The CTD measurements are stored in the CTD's internal memory and are also transmitted to the Aquadopp over the serial connection, where they are stored together with the DF7 current data in the [filename]_avgd.aqd file.

Note: It is not possible to communicate with the Aquadopp while the CTD is connected, as the CTD occupies the 8-pin connector used for communication with the instrument.

Configuring the SBE 37 in Nortek Deployment

Before configuring the Aquadopp for the deployment, configure and start the SBE 37 using Nortek Deployment.

  1. Open Nortek Deployment and navigate to the External CTD tab.

    External CTD menu
  2. Connect the SBE 37 to your computer using the cable provided by Sea-Bird.
  3. Under Select COM port for CTD sensor configuration, select the COM port associated with the SBE 37. If you need help identifying the correct COM port, refer to Connection problems.
  4. Under Select CTD sensor type, select 37-SM.
  5. Click Connect.
  6. Once connected, the upper part of the External CTD screen displays the live data output from the SBE 37. This can be used to verify communication and troubleshoot the CTD.

    External CTD live view
  7. Click Set Nortek parameters to configure the SBE 37 with the parameters required for integration with the Nortek instrument.
  8. If required, click Clear CTD memory to clear data stored in the internal memory of the SBE 37.
  9. When you are ready to deploy the CTD, click Start. The SBE 37 will start recording and automatically disconnect from Nortek Deployment. A confirmation message is displayed when the CTD has started successfully.

    External CTD start confirmation

Configuring the Aquadopp

Once the SBE 37 has been configured and started, configure the Aquadopp for the deployment.

  1. Connect to your instrument in Nortek Deployment. 
  2. Under Plan and Deploy, you will find the option for External sensor under the Hardware step. Enable External sensor and select SBE 37 as the external sensor. The Power output is set automatically to match the sensor you choose and should not be altered.

    Hardware tab of the Plan & Deploy menu
  3. Configure the remaining deployment settings as required.
  4. Start the deployment.
  5. Connect the SBE 37 to the instrument's 8-pin connector. The SBE 37 will transmit its measurements to the instrument over the serial connection.

Requirements and Communication Settings

To connect to the CTD, the following requirements must be met:

  • Firmware version 3.0 or higher

  • Baud rate set to 9600

Use the Sea-Bird software to update the firmware and set the correct baud rate.