Official comment Hi Marcel, Thank you for posting to our forum, and I am sorry to hear you have problems with the download speed. Unfortunately, since the Vector and HR instruments produce a significant amount of ...
-
Hi Ma'ruf, Unfortunately, there is no way to say exactly what the remaining battery capacity is. The voltage of the alkaline batteries falls quickly at the beginning, slowly during most of its life...
-
In the program Signature Deployment, you can connect to the instrument via the terminal and send the following commands: GETUSER SETUSER,DECL=XX SAVE,USER where XX is the desired compass offset. Th...
-
The first sample in every burst in diagnostic mode is done without a transmitted signal. The low amplitude seen at the beginning of each burst is therefore the noise floor measurement. The values s...
-
Try to reformat the recorder. This should always be done after upgrading the firmware in any Signature instrument. To do so, first right click on the instrument in the discover screen and select we...
-
Even though the instrument is set up to measure in XYZ-coordinates, we unfortunately only have functionality for outputting processed wave data from the prolog with directions relative to ENU-coord...